SENSOR & TRANSMITTER

UT-9003

  • PRESSURE & VACUUM TRANSMITTER - MK

     

    ♦ Long term stability batter then ±0.1 % fs/yr 0.25 % Full span Accuracy.

    ♦ Fully ss 316 body.

    ♦ Choice of ma V or radiometric output.

    ♦ Accuracy:0.25% on FSD.

    ♦ Thin fill technology for long time stability.

    ♦ range:0-10,16,40,50,60,100,160,200,300,400,700,1000 bar.

    ♦ Vacuum range:-1 to 0,-1 to 1,-1 to 3,-1 to 5 bar.

  •  FEATURE

     

    ♦ Pressure range: 1 bar to 50 bar | 15 psi to 700 psi (absolute and sealed gage)

    ♦  Ratiometric output: 0.5 Vdc to 4.5 Vdc or 0.33 Vdc to 2.97 Vdc

    ♦ 4 mA to 20 mA output supports transmission of output signals over longer distances

    ♦  Calibrated and temperature compensated

    ♦ Total Error Band: ±1.0 %FSS from -20°C to 85°C [-4°F to 185°F]

    ♦ External freeze/thaw resistance: 6 cycles from -30°C to 50°C [-22°F to 122°F] • Insulation resistance: >100 MOhm, 1500 Vdc (in dry, nonionized air)

    ♦  Dielectric strength: AC1500V, 1 min. or AC1800V, 1 s (in dry, nonionized air)

    ♦ Current consumption: 3.5 mA max.

    ♦ EMC (radiated immunity): 200 V/m (ratiometric voltage output), 140 V/m (current output) per ISO 11452-2

    ♦  Ingress protection IP67 (Metri-Pack connector), IP69K (cable harness), IP65 (DIN connector)

    ♦ RoHS, REACH, and CE compliant

    ♦ Six industry-standard pressure port types, including a tube port which provides for hermetically-sealed process connection